Contact Sales/Support: +1 (888) 800-9580
정기 유지보수: 당사의 사이트와 서비스가 8월 16일 오후 8:30분부터 오후 10:30분(EDT)까지 간헐적으로 중단될 수 있습니다. 양해해 주셔서 감사합니다.


i7 Reflection Port During Transmission Measurement

Q - Should there be anything in the reflection view port when taking transmission measurements?

A - Yes.  The Color i7 transmission kit includes a transmission port plate and a white spectralon disk to be used for transmission calibrations and measurements.  The transmission port plate is to be used when doing transmission measurements other than Haze or Reflectance/Transmission.  When measuring Haze or Reflectance/Transmission the LAV port plate is used with the transmission spectralon disk at the reflectance port.


Need a Quote? Contact Sales(888) 800-9580

Technical Questions? Contact Support(888) 826-3042